Related Experiment Video
Updated: Jun 19, 2026

Rejection of Fluorescence Background in Resonance and Spontaneous Raman Microspectroscopy
Published on: May 18, 2011
Ultrashort-pulse measurement using noninstantaneous nonlinearities: Raman effects in frequency-resolved optical
Abstract:
Ultrashort-pulse-characterization techniques generally require instantaneously responding media. We show that this is not the case for frequency-resolved optical gating (FROG). We include, as an example, the noninstantaneous Raman response of fused silica, which can cause errors in the retrieved pulse width of as much as 8% for a 25-fs pulse in polarization-gate FROG. We present a modified pulse-retrieval algorithm that deconvolves such slow effects and use it to retrieve pulses of any width. In experiments with 45-fs pulses this algorithm achieved better convergence and yielded a shorter pulse than previous FROG algorithms.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...