Related Experiment Video
Updated: Jun 19, 2026

In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
Heparin concentration-based anticoagulation for cardiac surgery fails to reliably predict heparin bolus dose
Sean Garvin1, Daniel C FitzGerald, George Despotis
1Brigham and Women's Hospital, Harvard Medical School, Boston, MA 02215, USA.
Background:
Hemostasis management has evolved to include sophisticated point-of-care systems that provide individualized dosing through heparin concentration-based anticoagulation. The Hepcon HMS Plus system (Medtronic, Minneapolis, MN) estimates heparin dose, activated clotting time (ACT), and heparin dose response (HDR). However, the accuracy of this test has not been systematically evaluated in large cohorts.
Methods:
We examined institutional databases for all patients who underwent cardiac surgery with cardiopulmonary bypass (CPB) at our institution from February 2005 to July 2008. During this period, the Hepcon HMS Plus was used exclusively for assessment of heparin dosing and coagulation monitoring. Detailed demographic, surgical, laboratory, and heparin dosing data were recorded. ACT, calculated and measured HDR, and heparin concentrations were recorded. Performance of the Hepcon HMS Plus was assessed by comparison of actual and target ACT values and calculated and measured HDR.
Results:
In 3880 patients undergoing cardiac surgery, heparin bolus dosing to a target ACT resulted in wide variation in the postheparin ACT (r(2) = 0.03). The postheparin ACT did not reach the target ACT threshold in 7.4%(i.e., when target ACT was 300 s) and 16.9% (i.e., when target ACT was 350 s) of patients. Similarly, the target heparin level calculated from the HDR did not correlate with the postbolus heparin level, with 18.5% of samples differing by more than 2 levels of the assay. Calculated and measured HDR were not linearly related at any heparin level.
Conclusions:
The Hepcon HMS Plus system poorly estimates heparin bolus requirements in the pre-CPB period. Further prospective studies are needed to elucidate what constitutes adequate anticoagulation for CPB and how clinicians can reliably and practically assess anticoagulation in the operating room.
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Venous Thrombosis III: Interprofessional Care
Depolarizing Blockers: Pharmocokinetics
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses