Related Experiment Video
Updated: Jun 19, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Efficient use of automatic exposure control systems in computed tomography requires correct patient positioning
J Gudjonsdottir1, J R Svensson, S Campling
1Röntgen Domus Medica, Reykjavík, Iceland. joninag@hi.is
Background:
Image quality and radiation dose to the patient are important factors in computed tomography (CT). To provide constant image quality, tube current modulation (TCM) performed by automatic exposure control (AEC) adjusts the tube current to the patient's size and shape.
Purpose:
To evaluate the effects of patient centering on tube current-time product (mAs) and image noise.
Material And Methods:
An oval-shaped acrylic phantom was scanned in various off-center positions, at 30-mm intervals within a 500-mm field of view, using three different CT scanners. Acquisition parameters were similar to routine abdomen examinations at each site. The mAs was recorded and noise measured in the images. The correlation of mAs and noise with position was calculated using Pearson correlation.
Results:
In all three scanners, the mAs delivered by the AEC changed with y-position of the phantom (P<0.001), with correlation values of 0.98 for scanners A and B and -0.98 for scanner C. With x-position, mAs changes were 4.9% or less. As the phantom moved into the y-positions, compared with the iso-center, the mAs varied by up to +70%, -34%, and +56% in scanners A, B, and C, respectively. For scanners A and B, noise in two regions of interest in the lower part of the phantom decreased with elevation, with correlation factors from -0.95 to -0.86 (P<0.02). In the x-direction, significant noise relationships (P<0.005) were only seen in scanner A.
Conclusion:
This study demonstrates that patient centering markedly affects the efficacy of AEC function and that tube current changes vary between scanners. Tube position when acquiring the scout projection radiograph is decisive for the direction of the mAs change. Off-center patient positions cause errors in tube current modulation that can outweigh the dose reduction gained by AEC use, and image quality is affected.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
06:50Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Anatomical Positions
The body is upright, facing forward, and standing erect.
The feet are parallel and flat on the floor.
The arms are hanging by the...