Related Experiment Video
Updated: Jun 19, 2026

Analysis of 18FDG PET/CT Imaging as a Tool for Studying Mycobacterium tuberculosis Infection and Treatment in Non-human Primates
Published on: September 5, 2017
Development of a generic thresholding algorithm for the delineation of 18FDG-PET-positive tissue: application to the
S Vauclin1, K Doyeux, S Hapdey
1LITIS Laboratory EA 4108-QUANT.I.F, University of Rouen, Rouen, France. Siemens Medical Solutions, Bobigny, France.
Abstract:
An iterative generic algorithm has been developed to compare three thresholding models used to delineate gross tumour volume on (18)F-FDG PET images. 3D volume was extracted and characteristic parameters were measured. Three fitting models using different parameters were studied: model 1 (volume, contrast), model 2 (contrast) and model 3 (SUV). The calibration was performed using a cylindrical phantom filled with hot spheres. To validate the models, two other phantoms were used. The calibration procedure showed a better fitting model for model 1 (R(2) from 0.94 to 1.00) than for model 3 (0.95) and model 2 (0.69). The validation study shows that model 3 yielded large volume measurement errors. Models 1 and 2 gave close results with no significant differences. Model 2 was preferred because it presents less error dispersion and needs fewer characteristic parameters, making it easier to implement. Our results show the importance of developing a generic algorithm to compare the performances of fitting models objectively and to validate results on other phantoms than the ones used during the calibration process to avoid methodological biases.