A High-throughput-compatible FRET-based Platform for Identification and Characterization of Botulinum Neurotoxin Light Chain Modulators
Published on: December 27, 2013
Lab-on-a-chip for botulinum neurotoxin a (BoNT-A) activity analysis
Steven Sun1, Miguel Ossandon, Yordan Kostov
1Division of Biology, Office of Science and Engineering Laboratories, FDA, NIH/NCI, Silver Spring, MD 20993, USA.
Lab on a Chip
|October 30, 2009
Summary
A novel Lab-on-a-chip device detects botulinum neurotoxin serotype A (BoNT-A) activity by measuring peptide cleavage. This portable system offers sensitive detection for potential use in point-of-care diagnostics.
Area of Science:
- Biomedical Engineering
- Analytical Chemistry
- Toxicology
Background:
- Botulinum neurotoxin serotype A (BoNT-A) detection is crucial for diagnostics.
- Existing methods for BoNT-A detection can be complex and require laboratory settings.
- There is a need for rapid, sensitive, and portable detection systems.
Purpose of the Study:
- To design, fabricate, and test a Lab-on-a-chip (LOC) device for in vitro detection of BoNT-A activity.
- To develop a portable fluorescence detection system for BoNT-A.
- To assess the sensitivity and potential of the LOC system for point-of-care diagnostics.
Main Methods:
- Fabrication of a six-layer LOC device using Laminated Object Manufacturing (LOM) with Poly(methyl methacrylate) (PMMA) and Polycarbonate (PC) layers.
- Development of a Förster Resonance Energy Transfer (FRET) assay utilizing a fluorophore-tagged SNAP-25 peptide substrate specific for BoNT-A light chain (LcA).
- Measurement of FRET fluorescence using a portable charge-coupled device (CCD) fluorescent detector with multi-wavelength light-emitting diode (LED) illumination.
Main Results:
- The LOC system successfully detected LcA activity.
- The system achieved a limit of detection as low as 0.5 nM for LcA, consistent with the in vitro assay sensitivity.
- The portable CCD detector enabled sensitive FRET fluorescence measurement.
Conclusions:
- The developed LOC device combined with a portable CCD detector provides a sensitive and simple system for BoNT-A activity detection.
- This point-of-care system has the potential for use in complex medical assays outside of traditional laboratory environments.
- The technology may improve healthcare delivery, particularly for underserved populations.
Related Concept Videos
Botulism
Botulism is a life-threatening neuroparalytic condition caused by botulinum neurotoxin, which is produced by the bacterium Clostridium botulinum, a Gram-positive, spore-forming, obligate anaerobe.In adults, the toxin enters the body in different ways: in foodborne botulism, the preformed toxin is absorbed in the intestine. In wound botulism, spores grow in injured tissue and release the toxin into the blood. Infant botulism differs mechanistically from adult forms. In infants, botulism commonly...
Directly Acting Muscle Relaxants: Dantrolene and Botulinum Toxin
Directly acting muscle relaxants like dantrolene and botulinum toxin (BoNT) have distinct mechanisms and applications. Dantrolene, a hydantoin derivative, acts on the ryanodine receptor (RYR1) in skeletal muscle cells. RYR1 are calcium channels present at the sarcoplasmic reticulum membrane. In response to excitation, they release calcium ions from the sarcoplasmic reticulum to the cytosol. Calcium promotes actin-myosin-mediated contraction of muscles.
The binding of dantrolene to the RYR1...
The binding of dantrolene to the RYR1...