Related Experiment Video
Updated: Jun 19, 2026

Generation of Murine Cardiac Pacemaker Cell Aggregates Based on ES-Cell-Programming in Combination with Myh6-Promoter-Selection
Published on: February 17, 2015
The immediate effects of pacemaker-related electric remodelling on left ventricular function in patients with sick
Yu-Chen Wang1, Yen-Hung Lin, Yen-Bin Liu
1Division of Cardiology, Department of Internal Medicine, China Medical University Hospital, Taichung, Taiwan.
Aims:
The immediate effects of electric remodelling on the left ventricular (LV) function by dual-chamber pacemakers remain unknown. The purpose of our study was to assess the interaction between heart rates and right ventricular pacing (VP) on LV contractility and diastolic function.
Method And Results:
Twenty-five patients with dual-chamber pacemakers and sick sinus syndrome were evaluated. Echocardiographic examinations included standard and tissue-Doppler echocardiography at bilateral mitral annulus margins under either the intrinsic atrio-ventricular sequential conduction (ventricular sensing; VS) mode or right ventricular apical pacing (VP) mode. Under either mode, we accelerated the pacing rate at an increment of 15 b.p.m. step-by-step from 60 to 90/min. The tissue-Doppler echocardiography of mitral annulus showed that under the VS status, accelerating atrial pacing rate from 60 to 90 b.p.m. enhanced A'-wave velocity (P < or =0.002), whereas no significant change of LV ejection fraction (LVEF) and E'-wave velocity were noted. Under the VP status, acceleration of pacing rates exerted no effect on the LVEF, E'-, and A'-wave (P = NS). While shifting the pacemaker mode from VS to VP, the E'-wave velocity (P < or =0.002) and E'/A' ratio decreased significantly (P < or = 0.001). The A'-wave velocity also increased significantly during shifting to VP mode at 60 b.p.m. (P < or =0.004).
Conclusion:
At fixed pacing rates, shifting from VS to VP mode impaired LV diastolic function immediately with preserved LV contractility. The acceleration of heart rate impaired LV diastolic function under VS mode.
More Related Videos
09:20Microelectrode Array Recording of Sinoatrial Node Firing Rate to Identify Intrinsic Cardiac Pacemaking Defects in Mice
Published on: July 5, 2021
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Cardiomyopathy III: Hypertrophic Cardiomyopathy
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism, and...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Cardiomyopathy II: Dilated Cardiomyopathy
Mechanism of Cardiac Arrhythmias