Performance evaluation of a new blood glucose monitor that requires no coding: the OneTouch Vita System
Joanna K Young1, John M Ellison, Robert Marshall
1Clinical Affairs, LifeScan Scotland, Ltd., Inverness, Scotland.
Background:
Improvements to blood glucose monitoring systems aim to simplify the testing process, reduce or eliminate errors, and provide additional information for patients with diabetes. New systems must continue to demonstrate high-quality analytical performance. The new OneTouch Vita System (LifeScan, Inc., Milpitas, CA) offers a no-code testing process and proven technology found in the OneTouch Ultra System. Comparative studies were conducted with the new and established systems to evaluate their precision and accuracy.
Methods:
Within-run precision in blood, total precision with controls, and system accuracy were evaluated using three lots of OneTouch Vita Test Strips and one lot of OneTouch Ultra Test Strips. Accuracy was tested across a wide glucose range (38-520 mg/dl, 2.1-28.9 mmol/liter) using fingertip blood samples from 139 subjects. Reference plasma glucose values were obtained using the YSI 2300 STAT Plus Glucose & Lactate Analyzer (YSI Inc., Yellow Springs, OH). All studies were designed in accordance with requirements published by the International Organization for Standardization (ISO 15197).
Results:
Precision testing (within-run and total) with both systems produced coefficients of variation (CVs) of <5% for all sample types and glucose levels. Within-run precision testing with blood showed CVs of <3.1% and <4.7% for the OneTouch Vita and OneTouch Ultra Systems respectively. Total precision with control samples gave CVs of <3.0% and <3.6% for the two systems. Consensus error grid analysis showed equivalent clinical accuracy with 98.4% (821/834) and 98.2% (273/278) of results within zone A. Both systems met the ISO acceptability requirements for system accuracy.
Conclusion:
The OneTouch Vita System provides a simple no-code testing process with performance comparable to the OneTouch Ultra and OneTouch Ultra2 Systems.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)