Performance of the rebound, noncontact and Goldmann applanation tonometers in routine clinical practice
Jose M Martinez-de-la-Casa1, Maria Jimenez-Santos, Federico Saenz-Frances
1Hospital Clinico San Carlos, Instituto de Investigaciones Oftalmologicas Ramon Castroviejo, Universidad Complutense, Madrid, Spain. martinezcasa@ya.com
Purpose:
To compare rebound tonometry (RBT) and noncontact tonometry (NCT) using Goldmann applanation tonometry (GAT) as reference.
Methods:
The study sample was comprised of 108 eyes of 108 subjects consecutively examined at a general ophthalmology clinic. The order of use of the three tonometers was randomized at the study outset. The difference between the methods was plotted against the mean to compare the tonometers. The hypothesis of zero bias was examined by a paired t-test and 95% limits of agreement (LoA) were also calculated. Differences with respect to GAT were assessed according to the international standard for ocular tonometers (ISO 8612).
Results:
Mean intraocular pressures (IOPs ± SD) obtained using the three instruments were GAT 17.5 ± 3.8 mmHg; RBT 18.5 ± 5.5 mmHg and NCT 17.4 ± 5.6 mmHg. The 95% LoA were from -7.9 to +7.7 mmHg for NCT-GAT and from -6.8 mmHg to +8.7 mmHg for RBT-GAT. A difference with respect to GAT under ± 1 mmHg was observed in 11.1% of the eyes measured by NCT and 18.5% of eyes measured by RBT. According to the IOP ranges established by the ISO 8612, differences from GAT measurements greater than ± 5 mmHg were always above the accepted level of 5%. Correlations between IOP and central corneal thickness (CCT) were significant for all three tonometers.
Conclusions:
The rebound and noncontact tonometer behaved similarly when used to measure IOP taking GAT measurements as the reference standard. Neither tonometer fulfilled ISO 8612 requirements. Both were similarly influenced by CCT.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...