Variations in arterial blood pressure are associated with parallel changes in FlowTrac/Vigileo-derived cardiac output
Savvas Eleftheriadis1, Zisis Galatoudis, Vasilios Didilis
1Department of Anesthesiology, University General Hospital of Alexandropoulis, Dragana, Alexandropoulis, PC 68100, Greece. sav.elef@yahoo.gr
Introduction:
The reliability of autocalibrated pressure waveform analysis by the FloTrac-Vigileo(R) (FTV) system for the determination of cardiac output in comparison with intermittent pulmonary arterial thermodilution (IPATD) is controversial. The present prospective comparison study was designed to determine the effects of variations in arterial blood pressure on the reliability of the FTV system in patients undergoing coronary artery bypass grafting (CABG).
Methods:
Comparative measurements of cardiac output by FTV (derived from a femoral arterial line; software version 1.14) and IPATD were performed in 16 patients undergoing elective CABG in the period before institution of cardiopulmonary bypass. Measurements were performed after induction of anesthesia, after sternotomy, and during five time points during graft preparation. During graft preparation, arterial blood pressure was increased stepwise in intervals of 10 to 15 minutes by infusion of noradrenaline and lowered thereafter to baseline levels.
Results:
Mean arterial blood pressure was varied between 85 mmHg and 115 mmHg. IPATD cardiac output did not show significant changes during periods with increased arterial pressure either during sternotomy or after pharmacological manipulation. In contrast, FTV cardiac output paralleled changes in arterial blood pressure; i.e. increased significantly if blood pressure was raised and decreased upon return to baseline levels. Mean arterial blood pressure (MAP) and FTV cardiac output were closely correlated (r = 0.63 (95% confidence interval [CI]: 0.49 - 0.74), P < 0.0001) while no correlation between MAP and IPATD cardiac output was observed. Bland-Altman analyses for FTV versus IPATD cardiac output measurements revealed a bias of 0.4 l/min (8.5%) and limits of agreement from 2.1 to -1.3 l/min (42.2 to -25.3%).
Conclusions:
Acute variations in arterial blood pressure alter the reliability of the FlowTrac/Vigileo device with the second-generation software. This finding may help to explain the variable results of studies comparing the FTV system with other cardiac output monitoring techniques, questions the usefulness of this device for hemodynamic monitoring of patients undergoing rapid changes in arterial blood pressure, and should be kept in mind when using vasopressors during FTV-guided hemodynamic optimization.
Related Concept Videos
Measurement of Blood Pressure
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation.
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...