Non-invasive cardiac output measurements based on bioreactance for optimization of atrio- and interventricular delays
Fakhar Z Khan1, Munmohan S Virdee, Peter J Pugh
1Addenbrooke's Hospital, Hills Road, Cambridge, UK.
Aims:
Non-invasive cardiac output monitoring (NICOM) based on bio-reactance offers a portable method to assess ventricular function. Optimization of cardiac resynchronization therapy (CRT) by echocardiography is labour-intensive. We compared the ability of NICOM and echocardiography to facilitate optimum CRT device programming.
Methods And Results:
Forty-seven patients in sinus rhythm were evaluated within 14 days of CRT implantation. The atrio- (AV) and interventricular (VV) delay intervals were incrementally adjusted and at each setting, NICOM and echocardiographic data were recorded. Left ventricular (LV) volumes and function were assessed by echocardiography at baseline and 3 months. Response to CRT was defined as a reduction in LV end-systolic volume (LVESV) by >15%. In all patients, cardiac output (CO) increased significantly at optimized settings compared with baseline (5.66 +/- 1.4 vs. 4.35 +/- 1.1 L/min, P < 0.001). A 20% increase in acute CO following CRT predicted LVESV reduction of >15% with a sensitivity of 81% and specificity of 92% (AUC 0.86). The optimum AV delay determined by NICOM was confirmed by echocardiography in 40 of 47 patients (85%, r = 0.89, P < 0.01) and for VV delay in 39 of 47 patients (83%, r = 0.89, P < 0.01).
Conclusion:
Non-invasive cardiac output monitoring is a simple, reliable, and portable alternative to echocardiography to program CRT devices.
Related Concept Videos
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...