Pressures available for transtracheal jet ventilation from anesthesia machines and wall-mounted oxygen flowmeters
Jens Fassl1, Uta Jenny, Sergei Nikiforov
1Simulation Development and Cognitive Science Laboratory, Pennsylvania State University College of Medicine, 500 University Ave., Hershey, PA 17033, USA.
Background:
Oxygen supplies capable of supporting transtracheal jet ventilators can be lifesaving. There is not much information about which oxygen sources (readily available inside and outside operating rooms) have sufficient driving pressure for transtracheal jet ventilation.
Methods:
We measured driving pressures (upstream or residual oxygen pressure) in a specially designed jet ventilation test system with a 2.25-mm (14-gauge) i.v. catheter. High-pressure oxygen sources evaluated included wall-mounted (Puritan, Allied Health, Precision, and Datex-Ohmeda) and anesthesia machine auxiliary oxygen flowmeters and oxygen flush valves from anesthesia machines (Draeger Narkomed 2B, Narkomed 4, Datex-Ohmeda Excel, and Datex-Ohmeda Modulus).
Results:
All 4 types of wall-mounted oxygen flowmeters, opened past their highest scale settings (15 L/min), delivered sufficient working pressures (range, 103-282 kPa; 16-41 psi). Working pressures from auxiliary oxygen flowmeters mounted on Datex-Ohmeda machines were adequate to support jet ventilation (range, 189-248 kPa; 27-36 psi), whereas those on tested Draeger machines did not supply sufficient pressure for jet ventilation: Narkomed 2B, 14-28 kPa (2-4 psi); Narkomed 4, 24-28 kPa (3-4 psi). Working pressures delivered by oxygen flush valves on tested Draeger machines were adequate to support jet ventilation, ranging from 96 to 117 kPa (14-17 psi), whereas pressures generated by tested Datex-Ohmeda flush valves were not (ranging from 50 to 62 kPa, 7-9 psi).
Conclusion:
Oxygen sources other than dedicated jet ventilator connectors to high-pressure pipeline oxygen may supply adequate working pressure, but each type of oxygen source needs testing to ensure that it supplies adequate working pressure.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings