Distance from file tip to the major apical foramen in relation to the numeric meter reading on the display of three
R A Higa1, C G Adorno, A K Ebrahim
1Pulp Biology and Endodontics, Department of Restorative Sciences, Graduate School, Tokyo Medical and Dental University, Tokyo, Japan. romikawa@hotmail.com
Aim:
To establish and compare the relationship between the distance from the file tip to the apical foramen and the numeric meter reading on the display of three different electronic apex locators (EALs).
Methodology:
A total of 12 extracted intact, straight, single-rooted human teeth with complete roots were used. The actual root canal length (AL) was determined after access preparation. For the electronic measurements with each EAL, silicon stops were fixed with auto-polymerizing resin to size 15 K-files at AL and 0.5, 1, 2, 3, 4 mm short of AL. The data was analysed by two-way anova and Tukey's honestly significant difference (HSD) test for multiple comparisons amongst EALs. Additionally, one-way anova and Tukey's HSD test were carried out for multiple comparisons amongst the measurements of each EAL.
Results:
There was a statistically significant difference amongst all EALs in indicating the position of file tips in relation to the major foramen (P < 0.05). The correlation between the meter reading and the position of the file tip from the apical foramen was statistically significant in the three EALs. There were significant differences amongst the measurements at distances from 0 to 2 mm in Justy III. In Dentaport, significant differences were found from 0 to 1 mm. However, the E-Magic Finder showed significant differences from 0 to 0.5 mm.
Conclusions:
Justy III was more capable of displaying the intracanal position of the file tip from the major foramen in mm whilst advancing through the root canal during electronic measurements than the Dentaport and E-Magic Finder Deluxe.
Related Concept Videos
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Adjusting a Traverse
Differential Leveling
Electronic Distance Measuring Instruments
Meridians