Accuracy of automatic couch corrections with on-line volumetric imaging
Winnie Li1, Douglas J Moseley1,2, Tony Manfredi3
1Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario, Canada.
Abstract:
The purpose of this study was to characterize automatic remote couch adjustment and to assess the accuracy of automatic couch corrections following localization with cone-beam CT (CBCT). Automatic couch movement was evaluated through passive reflector markers placed on a phantom, tracked with an optical tracking system (OTS). Repeated couch movements in the lateral, cranial/caudal, and vertical directions were monitored through the OTS to assess velocity and response time. In conjunction with CBCT, remote table movement for patient displacements following initial setup was available on four linear accelerators (Elekta Synergy). After the initial CBCT scan assessment, patients with isocenter displacements that exceeded clinical protocol tolerances were corrected using remote automatic couch movement. A verification CBCT scan was acquired after any remote movements. These verification CBCT datasets were assessed for the following time periods: one month post clinical installation, and six months later to monitor remote couch correction stability. Residual error analysis was evaluated using the verification scans. The mean +/- standard deviations (mu +/- sigma) of couch movement based on phantom measurements with the OTS were 0.16 +/- 0.48 mm, 0.32 +/- 0.30 mm, 0.11 +/- 0.12 mm in the L/R, A/P, and S/I couch directions, respectively. The fastest maximum velocity was observed in the inferior direction at 10.5 mm/s, and the slowest maximum velocity in the left direction at 3.6 mm/s. From 1134 verification CBCT registrations for 207 patients, the residual error for each translational direction from each month evaluated are reported. The mu was less than 0.3 mm in all directions, and sigma was in the order of 1 mm. At a 3 mm threshold, 21 of the 1134 fractions (2%) exceeded tolerance, attributed to patient intrafraction movement. Remote automatic couch movement is reliable and effective for adjusting patient position with a precision of approximately 1mm. Patient residual error observed in this study demonstrates that displacement is minimal after remote couch adjustment.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies VI: Voiding Cystourethrography and Cystography