Physical and clinical comparison between a screen-film system and a dual-side reading mammography-dedicated computed
S Rivetti1, B Canossi, R Battista
1Department of Health Physics, AUSL Modena, Modena, Italy.
Background:
Digital mammography systems, thanks to a physical performance better than conventional screen-film units, have the potential of reducing the dose to patients, without decreasing the diagnostic accuracy.
Purpose:
To achieve a physical and clinical comparison between two systems: a screen-film plate and a dual-side computed radiography system (CRM; FUJIFILM FCR 5000 MA).
Material And Methods:
A unique feature of the FCR 5000 MA system is that it has a clear support medium, allowing light emitted during the scanning process to be detected on the "back" of the storage phosphor plate, considerably improving the system's efficiency. The system's physical performance was tested by means of a quantitative analysis, with calculation of the modulation transfer function, detective quantum efficiency, and contrast-detail analysis; subsequently, the results were compared with those achieved using a screen-film system (SFM; Eastmann Kodak MinR-MinR 2000). A receiver operating characteristic (ROC) analysis was then performed on 120 paired clinical images obtained in a craniocaudal projection with the conventional SFM system under standard exposure conditions and also with the CRM system working with a dose reduced by 35% (average breast thickness: 4.3 cm; mean glandular dose: 1.45 mGy). CRM clinical images were interpreted both in hard copy and in soft copy.
Results:
The ROC analysis revealed that the performances of the two systems (SFM and CRM with reduced dose) were similar (P>0.05): the diagnostic accuracy of the two systems, when valued in terms of the area underneath the ROC curve, was found to be 0.74 for the SFM, 0.78 for the CRM (hard copy), and 0.79 for the CRM (soft copy).
Conclusion:
The outcome obtained from our experiments shows that the use of the dual-side CRM system is a very good alternative to the screen-film system.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT