The Flex-It stylet is less effective than a malleable stylet for orotracheal intubation using the GlideScope
Timothy P Turkstra1, Philip M Jones, Katherine M Ower
1Department of Anesthesia and Perioperative Medicine, London Health Sciences Centre, University Hospital, Room C3-104, 339 Windermere Rd., London, Ontario, Canada N6A 5A5. timothy.turkstra@londonhospitals.ca
Background:
The GlideScope videolaryngoscope (Verathon Medical, Bothell, WA) usually provides excellent glottic visualization, but directing an endotracheal tube through the vocal cords can be challenging. The goal of the study was to compare the dedicated Flex-It stylet (FIS, Parker Medical, Highlands Ranch, CO) with a malleable stylet, as assessed by time to intubation (TTI).
Methods:
Eighty patients requiring orotracheal intubation for elective surgery were randomly allocated to either the FIS or a malleable stylet (control) to facilitate tracheal intubation using the GlideScope. TTI was recorded by blinded assessors; operators were blinded until after laryngoscopy. The operator assessed the ease of intubation using a 100-mm visual analog scale (0 = easy to 100 = difficult). The number of intubation attempts, number of failures, glottic grades, and use of external laryngeal manipulation were documented.
Results:
The median TTI was 41 s (interquartile range [IQR] 30-51) for the Flex-It group compared with 32 s (IQR 28-42) for the control group (P = 0.03). The median visual analog scale score for ease of intubation was 20 (IQR 11-39) for the Flex-It group compared with 15 (IQR 8-28) for the control group (P = 0.13). The overall incidence of a Cormack-Lehane Grade I or II glottic view was 100%.
Conclusions:
In a group of experienced operators using the GlideScope, the FIS was less effective for orotracheal intubation than a malleable endotracheal tube stylet.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Suctioning the Nasopharyngeal Airway
Equipment Required
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...