Manual ventilation devices in neonatal resuscitation: tidal volume and positive pressure-provision
Charles C Roehr1, Marcus Kelm, Hendrik S Fischer
1Clinic of Neonatology, Charité Universitätsmedizin Berlin, Charitéplatz 1, 10117 Berlin, Germany. christoph.roehr@charite.de
Background:
Excessive peak inspiratory pressures (PIP) and high tidal volumes (Vt) during manual ventilation can be detrimental to the neonatal lung. We compared the influence of different manual ventilation devices and individual professional experience on the extent of applied Vt and PIP in simulated neonatal resuscitation.
Material And Methods:
One hundred and twenty medical professionals were studied. An intubated mannequin (equivalent to 1.0 kg neonate) was ventilated using two different devices: a self-inflating bag and a T-piece resuscitator. Target value was a PIP of 20 cm H(2)O. Applied PIP and the resulting Vt were recorded continuously using a respiratory function monitor (CO(2)SMO(+), Novametrix, USA).
Results:
Vt and PIP provision was significantly higher in SI-bags, compared to T-piece devices: median (interquartile range) PIP 25.6 (18.2) cm H(2)O vs 19.7 (3.2) cm H(2)O (p<0.0005), and Vt 5.1(3.2) ml vs Vt 3.6 (0.8) ml (p<0.0005) respectively. The intersubject variability of Vt and PIP provision was distinctly higher in SI-bags, compared to T-piece devices. Professional experience had no significant impact on the level and the variability of Vt or PIP provided.
Conclusion:
Use of T-piece devices guarantees reliable and constant Vt and PIP provision, irrespective of individual, operator dependent variables. Methods to measure and to avoid excessive tidal volumes in neonatal resuscitation need to be developed.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...