Related Experiment Video
Updated: Jun 18, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Radiation therapy photon beams dose conformation according to dose distribution around intracavitary-applied
Slaven Jurković1, Gordana Zauhar, Dario Faj
1Department of Radiotherapy and Oncology, University Hospital, Rijeka, Croatia. slaven.jurkovic@ri.htnet.hr
Abstract:
Intracavitary application of brachytherapy sources followed by external beam radiation is essential for the local treatment of carcinoma of the cervix. Due to very high doses to the central portion of the target volume delivered by brachytherapy sources, this part of the target volume must be shielded while being irradiated by photon beams. Several shielding techniques are available, from rectangular block and standard cervix wedge to more precise, customized step wedge filters. Because the calculation of a step wedge filter's shape was usually based on effective attenuation coefficient, an approach that accounts, in a more precise way, for the scattered radiation, is suggested. The method was verified under simulated clinical conditions using film dosimetry. Measured data for various compensators were compared to the numerically determined sum of the dose distribution around brachytherapy sources and one of compensated beam. Improvements in total dose distribution are demonstrated, using our method. Agreement between calculation and measurements were within 3%. Sensitivity of the method on sources displacement during treatment has also been investigated.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Radiation: Applications
The average...