[Evaluation of flow rate regulators for intravenous infusion]
T Caruba1, L Havard, F Gillaizeau
1Service de pharmacie, hôpital européen Georges-Pompidou, Assistance publique-Hôpitaux de Paris, 20, rue Leblanc, 75015 Paris, France. thibaut.caruba@egp.aphp.fr
Objective:
The aim of our study is to evaluate seven flow rate regulators (FRR) to assess the reliability of these devices compared to standard perfuser with roller clamp.
Study Design:
Each FRR was tested with 5% dextrose and 0.9% sodium chloride combined with three different theoretical flow rates (30, 80 and 250 ml/h). Accuracy was compared with the theoretical value. Repeatability of flow rate was assessed thanks to variance break-up.
Results:
Each FFR exhibits at least one combination "flow rate-solution" significantly different of the theoretical flow rate. Exadrop was the least successful of the FFR according to the accuracy. This FFR had for each combination a flow rate different of the theoretical (mean error: -24.0 ml/h). Tutodrop was the most successful of the FFR according to the accuracy with five combinations comparable to the theoretical value (mean error: -1.2 ml/h). The standard perfuser with roller clamp, used without FRR, reported two combinations comparable to the theoretical value and showed lowest rates for repeatability.
Conclusion:
Our study exhibits the poor performances of the FRR studied: according to expected flow regulation, the reported results demonstrate the lack of accuracy. Their only one value added compare to the roller clamp is to improve the repeatability of the flow rate.
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Pipe Flowrate Measurement
The orifice meter is a simple,...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
IV Infusion to Oral Dosing: Conversion Methods
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations