Use of bioabsorbable plates for cranial fixation
Kosumo Noda1, Rokuya Tanikawa, Toshihide Sugimura
1Dohtoh Neurosurgical Hospital, Kitami, Hokkaido.
Neurologia Medico-Chirurgica
|November 27, 2009
Summary
Bioabsorbable LactoSorb plates offer a strong alternative to titanium for cranial fixation. Inserting these plates into the diploe overcomes thickness and cost issues, yielding favorable aesthetic results.
Area of Science:
- Biomaterials Science
- Neurosurgery
- Plastic Surgery
Background:
- Titanium plates are standard for cranial fixation but cause aesthetic concerns and radiographic artifacts.
- Bioabsorbable plates address these issues but can be thicker and more expensive.
- LactoSorb, a bioabsorbable polymer, demonstrates strength comparable to titanium.
Purpose of the Study:
- To present a novel technique for cranial fixation using bioabsorbable LactoSorb plates.
- To overcome the disadvantages of thickness and cost associated with absorbable plates.
- To evaluate the aesthetic and safety outcomes of this technique in adult patients.
Main Methods:
- Developed a technique of inserting LactoSorb plates into the diploe for cranial fixation.
- Applied the method in a cohort of 46 patients undergoing craniotomy.
- Utilized a bioabsorbable polymer composed of 82% poly-L-lactic acid and 18% polyglycolic acid.
Main Results:
- Aesthetically favorable outcomes were achieved in all 46 patients.
- No intraoperative or postoperative complications were reported.
- The technique effectively addressed the limitations of standard absorbable plates.
Conclusions:
- Inserting bioabsorbable plates into the diploe is a viable and effective method for cranial fixation.
- This technique offers aesthetic advantages and avoids radiographic artifacts associated with titanium.
- Bioabsorbable plates, particularly LactoSorb, show potential to replace titanium for cranial fixation, especially in adults with thin scalps.