Optimisation of trapping parameters in preparative capillary gas chromatography for the application in
Cornelia Meinert1, Werner Brack
1UFZ - Helmholtz Centre for Environmental Research, Department of Effect-Directed Analysis, Permoserstrasse 15, D-04318 Leipzig, Germany. cornelia.meinert@ufz.de
Abstract:
Preparative capillary gas chromatography (pcGC) provides novel high resolution fractionation opportunities in effect-directed analysis. However, harvesting efficiency strongly depends on the operating parameters of the system. Therefore, the performance of the pcGC system was optimised by identifying the best operating parameters for the preparative fraction collector (PFC) using six test analytes with different physicochemical properties. The present study indicates that pcGC parameters need to be selected individually for the investigated analytes. The major focus was put on the trapping parameters as published findings on optimum trapping conditions are very variable. No generally agreed concept is available. An alternative to temperature-controlled trapping are solvent-filled traps. The solvent dichloromethane (DCM) proved to be most suitable for a large range of compounds. Recoveries are equal to optimised dry trapping at defined temperature. Optimised recoveries were in the range of 50-70% for all compounds except benzo[a]pyrene with a recovery of 94% using one PFC and DCM-filled traps at trapping temperature of -10 degrees C, at PFC temperatures of 300 degrees C for phenol, 400 degrees C for benzo[a]pyrene and 320 degrees C for the remaining analytes.
More Related Videos
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
10:14Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...