Introduction:
Hypovolemia and hypovolemic shock are life-threatening conditions that occur in numerous clinical scenarios. Near-infrared spectroscopy (NIRS) has been widely explored, successfully and unsuccessfully, in an attempt to use it as an early detector of hypovolemia by measuring tissue oxygen saturation (StO2). In order to investigate the measurement site dependence and probe dependence of NIRS in response to hemodynamic changes, such as hypovolemia, we applied a simple cardiovascular challenge: a posture change from supine to upright, causing a decrease in stroke volume (as in hypovolemia) and a heart rate increase in combination with peripheral vasoconstriction to maintain adequate blood pressure.
Methods:
Multi-depth NIRS was used in nine healthy volunteers to assess changes in StO2 in the thenar and forearm in response to the hemodynamic changes associated with a posture change from supine to upright.
Results:
A posture change from supine to upright resulted in a significant increase (P < 0.001) in heart rate. Thenar StO2 did not respond to the hemodynamic changes following the posture change, whereas forearm StO2 did. Forearm StO2 was significantly lower (P < 0.001) in the upright position compared to supine for all probing depths.
Conclusions:
The primary findings in this study were that forearm StO2 is a more sensitive parameter to hemodynamic changes than thenar StO2 and that the depth at which StO2 is measured is of minor influence. Our data support the use of forearm StO2 as a sensitive parameter for the detection of central hypovolemia and hypovolemic shock in (trauma) patients.
More Related Videos
04:44Near-Infrared Spectroscopy During Reactive Hyperemia for the Assessment of Lower Limb Vascular Function
Published on: March 22, 2024
09:04Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.