Related Experiment Video
Updated: Jun 18, 2026

Conducting Respiratory Oscillometry in an Outpatient Setting
Published on: April 8, 2022
Non-contact respiratory rate measurement validation for hospitalized patients
Amy D Droitcour1, Todd B Seto, Byung-Kwon Park
1Kai Sensor, Honolulu, HI 96816, USA. adroitcour@kaisensors.com
Abstract:
This paper presents the first clinical results for validating the accuracy of respiratory rate obtained for hospitalized patients using a non-contact, low power 2.4 GHz Doppler radar system. Twenty-four patients were measured in this study. The respiratory rate accuracy was benchmarked against the respiratory rate obtained using Welch Allyn Propaq Encore model 242, the Embla Embletta system with Universal XactTrace respiratory effort sensor and Somnologica for Embletta software, and by counting chest excursions. The 95% limits of agreement between the Doppler radar and reference measurements fall within +/-5 breaths per minute.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Volumes and Capacities I
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Temperature Measurement Sites