Related Experiment Video
Updated: Jun 18, 2026

Functional Neuroimaging Using Ultrasonic Blood-brain Barrier Disruption and Manganese-enhanced MRI
Published on: July 12, 2012
Perturbation analysis of the effects of B1+ errors on parallel excitation in MRI
Chao Ma1, Dan Xu, Kevin F King
1Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, IL 61801, USA. chaoma2@illinois.edu
Abstract:
Design of RF pulses for parallel excitation using phased array transmit coils in MRI requires the B1+ maps that are estimated from B1+ mapping experiments. This paper characterizes the effects of B1+ mapping errors on the resulting excitation pattern using a small perturbation analysis based on linearization of the Bloch equation. The accuracy of the proposed perturbation analysis is validated by Bloch equation simulations based on experimental B1+ maps. The perturbation analysis builds a transparent connection between the B1+ mapping errors and the resulting excitation errors, and can be used to study the robustness of the designed RF pulses to B1+ mapping errors. The proposed method may also supply useful metrics for designing RF pulses that are robust to B1+ mapping errors.
More Related Videos
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
08:43Application of Granger Causality Analysis of the Directed Functional Connection in Alzheimer's Disease and Mild Cognitive Impairment
Published on: August 7, 2017
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Magnetic Resonance Imaging