The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Static forces variation and pressure distribution in laryngoscopy performed by straight and curved blades
S Cecchini1, S Silvestri, M Carassiti
1Faculty of Biomedical Engineering, University Campus Bio-Medico of Rome, Italy.
Abstract:
A theoretical analysis of the forces acting on the laryngoscope during the lifting of the epiglottis is carried out by applying the basic principles of statics. The static model of a laryngoscope equipped with a straight and a curved blade and the forces variation, as a function of the introduction angle and of tissue reaction application point, are described. The pharyngeal tissues and epiglottis pressure distribution on the blade is obtained, with a 1mm(2) resolution, by measurements performed in-vitro on a simulation mannequin, using straight and curved blades. The straight blade requires more effort than the curved one to obtain the same visualization of vocal cords, however forces exerted by using a laryngoscope with a curved blade do not vary linearly with the application point of tissue reaction. Average intensity of the tissue reaction has been found in the order of 32+/-11 N. Pressure distribution is maximally concentrated on the tip of curved blades (0.5 MPa on 5mm axial length), whereas it is more dispersed on straight blades (0.2 MPa on 10mm axial length). The inclination of the handle also influences the effort of the operator: for both blades, from 0 rad to 1.57 rad, the lifting force shows a total variation of about 13% of the top value, the transversal forces vary less than 6% of the top value.
Related Concept Videos
Hydrostatic Pressure Force on a Curved Surface
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Fluid Pressure over Curved Plate of Constant Width
Fluid Pressure over Flat Plate of Variable Width
The pressure distribution on the plate can be calculated by determining the force that acts on a differential area strip of the plate. Thus, the magnitude of the force is equal to the...
Static, Stagnation, Dynamic and Total Pressure
Suctioning the Nasopharyngeal Airway
Equipment Required