Related Experiment Video
Updated: Jun 18, 2026

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
Published on: June 1, 2012
Dissolved core alginate microspheres as "smart-tattoo" glucose sensors
Ayesha Chaudhary1, Monica Raina, Michael J McShane
1School of Biosciences and Bioengineering, Indian Institute of Technology Bombay, Powai, Mumbai, India. ayesha@iitb.ac.in
None:
The feasibility of multilayer thin film coated dissolved-core alginate-templated microsphere sensors based on fluorescence resonance energy transfer and competitive binding, was explored in simulated interstitial fluid, using glucose as a model analyte. The glucose sensitivity was observed to be 0.89%/mM glucose with a linear response in the range of 0-50 mM glucose. The sensor response was observed to be completely reversible in nature with a response time of 120 seconds. The system was further demonstrated to respond similarly using near-infrared dyes (Alexa Fluor-647-labeled dextran as donor and QSY-21-conjugated apo-GOx as acceptor) which exhibited a sensitivity of 0.94%/mM glucose with a linear response in range of 0-50 mM glucose, making the sensor more amenable to in vivo use, when implanted in scattering tissue.