Related Experiment Video
Updated: Jun 18, 2026

Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Dosimetric capabilities of the Iview GT portal imager using MCNP5 Monte Carlo simulations
1Chemical and Nuclear Engineering Department at the Polytechnic University of Valencia, Spain. bejusvi@iqn.upv.es
Abstract:
This work is focused on developing a methodology to obtain portal dosimetry with an amorphous silicon Electronic Portal Image Device (a-Si EPID) used in radiotherapy by means of Monte Carlo simulations and experimental measures. Pixel intensity values from portal images have been compared with dose measured from an ionization chamber and dose calculated from Monte Carlo simulations. To this end, several images were acquired with the Elekta iView GT EPID using an attenuator phantom slab (10 cm thickness of solid water) and a 6 MeV photon energy beam with different monitor units settings (MU). The average pixel value in a region of interest (ROI) centered at the beam for each image was extracted and compared to dose measures performed with an ionization chamber. These parameters were found to be linearly related with the number of monitor units. Since MCNP5 simulations allow calculating the deposited dose in the ROI within the phosphor layer of the EPID model, we could compare the portal dose with the simulated transit dose in order to perform a treatment control.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...