Comparison of three instruments for measuring red blood cell aggregation
Oguz K Baskurt1, Mehmet Uyuklu, Pinar Ulker
1Department of Physiology, Faculty of Medicine, Akdeniz University, Antalya, Turkey. baskurt@akdeniz.edu.tr
Clinical Hemorheology and Microcirculation
|December 10, 2009
Summary
Three instruments for measuring red blood cell (RBC) aggregation showed acceptable precision for detecting changes in plasma dilution. However, the RheoScan-A may not detect greatly enhanced aggregation, and the Myrenne M1 index may miss minor cell rigidity increases.
Area of Science:
- Hemorheology
- Biomedical Engineering
- Hematology
Background:
- Red blood cell (RBC) aggregation is a critical factor in hemorheology.
- Accurate measurement of RBC aggregation is essential for understanding various physiological and pathological conditions.
Purpose of the Study:
- To compare the performance of three instruments: LORCA, Myrenne Aggregometer, and RheoScan-A for measuring RBC aggregation.
- To assess the intra-assay reproducibility, biological variability, and sensitivity of these instruments under different conditions.
Main Methods:
- Workshop comparing LORCA, Myrenne Aggregometer, and RheoScan-A.
- Measurements included RBCs in diluted plasma, dextran, and glutaraldehyde-treated cells.
- Assessed intra-assay reproducibility, biological variability, and instrument sensitivity using standardized difference values.
Main Results:
- Intra-assay variations were generally below 5%, with exceptions for RheoScan-A amplitude (AMP) and surface area (SA).
- Biological variability was highest for half-time (T1/2), with other indices showing similar variability across devices.
- All instruments detected changes with plasma dilution; Myrenne and LORCA detected differences in dextran, but RheoScan-A did not.
- Glutaraldehyde treatment significantly affected parameters on all instruments, with varying sensitivity.
Conclusions:
- LORCA, Myrenne, and RheoScan-A demonstrate acceptable precision for detecting reduced RBC aggregation due to plasma dilution.
- The RheoScan-A may not detect significantly enhanced RBC aggregation, and the Myrenne M1 index might be insensitive to minor increases in cell rigidity.
- Further research is needed to define the optimal operating range for each instrument in RBC aggregation analysis.