Blood pressure monitoring: automated oscillometric devices
1Critikon, Inc., Tampa, FL 33631-3800.
Abstract:
An understanding of the principles of blood pressure measurement with automated oscillometric devices (e.g., Dinamap) allows a user to realize the purpose and level of accuracy for which the equipment is designed. Good technique cells for various actions: (1) using a cuff of proper size (too large a cuff will generally work acceptably, but too small a cuff will yield erroneously high readings); (2) squeezing all of the residual air out of the cuff before applying it to the arm or leg; (3) wrapping the cuff snugly around the arm or leg; (4) instructing the patient to refrain from talking or moving and letting nothing press against the cuff during measurement; (5) keeping the cuff and heart at the same horizontal level. (6) When instrument malfunction is suspected, first checking the status of the patient (not the status of the instrument) to be sure that a clinical emergency is not causing the suspected "malfunction"; (7) repeating a reading several times when it is suspicious and, if necessary, modifying the measurement conditions by using, for example, different limbs or different cuffs until the question is resolved; and (8) remembering that a leak in the cuff, hoses, or connection can cause an unexpected malfunction.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure