Background:
Stroke volume variation (SVV) is being increasingly used to predict fluid responsiveness. Since radial arterial pressure (RAP) and femoral arterial pressure (FAP) frequently showing discrepancies during liver transplantation (LT), we sought to investigate the effect of differing arterial waveforms on SVV and cardiac output (CO) derived from the Vigileo device, by comparing SVV and CO values derived from RAP (SVV(RAP), CO(RAP)) and FAP (SVV(FAP), CO(FAP)) during LT.
Methods:
The linear associations and agreements between SVV(RAP) and SVV(FAP) and between CO(RAP) and CO(FAP) were assessed during LT. Hemodynamic variables were measured at nine predefined time points in all 32 recipients, resulting in 288 data pairs.
Results:
Correlations were observed between SVV(RAP) and SVV(FAP) (r = .961) and between CO(RAP) and CO(FAP) (r = .848) at all time points. These correlations between SVV(RAP) and SVV(FAP) (r = .923) and between CO(RAP) and CO(FAP) (r = .902) existed even during the period when mean RAP and FAP values differed (10 minutes after reperfusion). Bland-Altman analysis for SVV(RAP) versus SVV(FAP) and for CO(RAP) versus CO(FAP) showed weak biases (-0.2% and -0.5 L/min) and reasonable limits of agreement (-2.2 to 1.8% and -1.9 to 0.9 L/min). The percentage errors for SVV and CO values were 27.0% and 22.2%.
Conclusions:
There was no significant difference between SVV(RAP) and SVV(FAP) when measured using the Vigileo device during LT. This finding indicated that SVV obtained using the Vigileo device offered relatively consistent information regardless of the catheterization site.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...