Abstract:
To assess the impact of bone cement viscosity on total knee arthroplasty, we compared 1 high-viscosity and 2 medium-viscosity cements with respect to mantle area and zone-specific intrusion depths into the tibial plateau. We analyzed postoperative radiographs to determine penetration area and depth in 72 consecutive patients (79 knees) in whom DePuy II (n = 11), Endurance (n = 34), or Simplex-P (n = 34) cement was used. Penetration into the tibial plateau (anteroposterior zones 1-4) was significantly reduced with use of the high-viscosity DePuy II cement but did not differ significantly between the 2 medium-viscosity cements, Endurance and Simplex-P. Surgical and tourniquet times were significantly decreased with the quicker setting DePuy II cement. Given these findings, additional studies are warranted to assess the long-term impact of the lower intrusion depths found with DePuy II cement. Such differences in cement penetration could jeopardize long-term fixation and lead to higher long-term device failure rates.