Related Experiment Video
Updated: Jun 17, 2026

Simultaneous Transcranial Alternating Current Stimulation and Functional Magnetic Resonance Imaging
Published on: June 5, 2017
Parallel excitation in the human brain at 9.4 T counteracting k-space errors with RF pulse design
Xiaoping Wu1, J Thomas Vaughan, Kâmil Uğurbil
1Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota Medical School, Minneapolis, Minnesota 55455, USA.
Abstract:
Multidimensional spatially selective radiofrequency (RF) pulses have been proposed as a method to mitigate transmit B1 inhomogeneity in MR experiments. These RF pulses, however, have been considered impractical for many years because they typically require very long RF pulse durations. The recent development of parallel excitation techniques makes it possible to design multidimensional RF pulses that are short enough for use in actual experiments. However, hardware and experimental imperfections can still severely alter the excitation patterns obtained with these accelerated pulses. In this note, we report at 9.4 T on a human eight-channel transmit system, substantial improvements in two-dimensional excitation pattern accuracy obtained when measuring k-space trajectories prior to parallel transmit RF pulse design (acceleration x4). Excitation patterns based on numerical simulations closely reproducing the experimental conditions were in good agreement with the experimental results.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
NMR Spectrometers: Resolution and Error Correction