Purpose:
To compare accelerated real-time two-dimensional (2D) and segmented three-dimensional (3D) cine steady-state free precession magnetic resonance (MR) imaging techniques by using a 32-channel coil with a conventional 2D cine imaging approach for imaging the heart and to evaluate any difference caused by free breathing and breath holding for real-time imaging.
Materials And Methods:
In this institutional review board-approved HIPAA-compliant study, 10 healthy volunteers and 22 consecutive patients who were suspected of having or were known to have heart disease underwent cardiac MR imaging by using a 32-channel coil. A conventional multisection 2D real-time cine sequence was used as the reference standard, and three additional accelerated cine sequences were implemented. Volumetric parameters, including ejection fraction (EF), end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume(SV), and myocardial mass, were derived. Wall motion and image quality were assessed by two radiologists. In addition, image time was registered. An additional set of images was acquired by using real-time sequences with free breathing, and quantitative measurements were compared with measurements on images obtained with breath holding. For quantitative analysis, repeated-measures analysis of variance, paired t test, and Bland-Altman analysis were used; for qualitative analysis, nonparametric Wilcoxon signed-rank test was used.
Results:
All volumetric measurements were significantly correlated with those of the standard sequence (r > 0.80, P < .01). No significant difference among protocols was observed in terms of mean levels for EF or ESV (P > .05). However, a significant difference was indicated for EDV and SV (P < .01).The accelerated protocols had significantly shorter image times (P < .001). Wall motion scores were concordant with the standard sequence in 43-44 (93%-96%) segments for the accelerated protocols, with a strong interreader agreement (intraclass correlation coefficient, > or =0.93). No significant difference was identified between real-time protocols with free breathing and those with breath holding for measurement of volumetric parameters.
Conclusion:
Accelerated real-time 2D and segmented 3D cine techniques are comparable to the standard clinical protocol in assessment of left ventricular global and regional parameters in substantially shorter image times.
More Related Videos
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Magnetic Resonance Imaging