Temperature dependence of gain shift in gamma-ray spectrometry system involving a long connecting cable
1National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Tsukuba, Ibaraki, Japan. kawada.t4@gmail.com
Abstract:
This work demonstrates the possibility of a gain shift or a peak-shape-deterioration with changes in the temperature of a long connecting cable between a pre-amplifier and a main amplifier in a Ge gamma-ray spectrometry system. The tests were performed for 50m-long RG174/U and RG58C/U cables with and without the termination. Such a temperature effect may cause from the temperature dependence of the resistivity of a central copper wire. In order to minimize such temperature effects in a Ge gamma-ray spectrometry, use of a terminator should be avoided.
More Related Videos
Related Concept Videos
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature from...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Thermal expansion and Thermal stress: Problem Solving
To solve the problem, first, identify the known and unknown quantities. The initial length (L) of the bridge is 1275 m, the coefficient of linear expansion (α) for steel is 12 x 10-6/°C, and the change in temperature (ΔT) is 55 °C.
Absorption of Radiation