Gold-coated pacemaker implantation after allergic reactions to pacemaker compounds
Thomas Syburra1, Ulrich Schurr, Mariette Rahn
1Cardiac Surgery Division, City Hospital Triemli Zurich, Birmensdorferstrasse 497, CH-8063 Zurich, Switzerland. thomas.syburra@triemli.stzh.ch
Pacemaker complications due to sterile wound necrosis resolved after switching to a gold-plated generator and polyurethane leads. Negative titanium allergy tests do not rule out pacemaker component hypersensitivity.
Area of Science:
- Cardiology
- Biomaterials Science
- Allergy and Immunology
Background:
- Pacemaker implantation is a common procedure for symptomatic atrio-ventricular block.
- Sterile wound necrosis is a rare but serious complication following pacemaker implantation.
- Identifying the cause of refractory wound complications can be challenging.
Observation:
- An 86-year-old male patient experienced recurrent sterile wound necrosis after pacemaker implantation for Mobitz II atrio-ventricular block.
- Despite negative microbiological tests and titanium skin patch testing, wound complications persisted, necessitating two generator re-implantations.
- A switch to a gold-plated generator and polyurethane leads resulted in complete resolution of wound complications.
Findings:
- The patient's recurrent sterile wound necrosis was successfully treated by replacing the pacemaker system with a gold-plated generator and polyurethane leads.
- This case suggests a potential hypersensitivity reaction to pacemaker components, even with negative standard allergy testing.
- Negative skin patch testing to titanium does not exclude an allergic reaction to pacemaker system components.
Implications:
- Gold-plated generators and polyurethane leads may be effective alternatives for patients with suspected pacemaker component allergies.
- Further investigation into hypersensitivity reactions to pacemaker materials is warranted.
- Clinicians should consider non-titanium components in refractory wound complications post-pacemaker implantation.