Balloon-tipped catheter for measuring urethral pressures
James S Walter1, John S Wheeler, Xiaoli Wang
1Hines VA Hospital (151), Research Service, Hines, IL 60141, USA. james.walter@va.gov
Background:
Better methods are needed for recording urethral function for complex urologic problems involving the bladder, urethra, and pelvic floor.
Objective:
To evaluate a balloon catheter for recording urethral pressure and function using bench-top testing and evaluation in an animal model.
Methods:
Balloon pressure-recording methods included slightly inflating the balloon with water and placing the pressure transducer on the distal end of the catheter. For bench-top testing, manual procedures and a silastic tube with a restriction were used. In 3 anesthetized dogs, pressure recorded from the skeletal urethral sphincter was induced with electrical stimulation of the sphincter. Anal sphincter pressure was also recorded.
Results:
Bench-top testing showed good pressure recordings, including a confined peak at the tube restriction. Animal tests showed urethral pressure records with rapid responses when electrical stimulation was applied. Peak pressure at the urethral skeletal sphincter was 55.7 +/- 15 cmH2O, which was significantly higher than the peak pressure recorded 2 cm distally in the proximal urethra (3.3 +/- 2.3 cmH2O). Peak anal pressures were smaller and unchanged for the 2 stimulations.
Conclusions:
Balloon-pressure recordings showed rapid responses that were adequate for the tests conducted. In the animal model, high-pressure contractions specific to the skeletal urethral sphincter were shown. Balloon-tipped catheters warrant further investigation and may have applications for the evaluation of detrusor-sphincter dyssynergia after spinal cord injury or for stress urinary incontinence.
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...