Related Experiment Video
Updated: Jun 17, 2026

09:17
In Vivo Luminal Measurement of Distension-Evoked Urothelial ATP Release in Rodents
Published on: September 7, 2022
Balloon-tipped catheter for measuring urethral pressures
James S Walter1, John S Wheeler, Xiaoli Wang
1Hines VA Hospital (151), Research Service, Hines, IL 60141, USA. james.walter@va.gov
The Journal of Spinal Cord Medicine
|December 23, 2009
Summary
A new balloon catheter effectively records urethral pressure and function. This method shows high-pressure contractions specific to the skeletal urethral sphincter, aiding in diagnosing urologic conditions.
Area of Science:
- Urology
- Biomedical Engineering
- Physiology
Background:
- Complex urologic issues require improved methods for recording urethral function.
- Current techniques may not adequately capture bladder, urethra, and pelvic floor dynamics.
Purpose of the Study:
- To assess a novel balloon catheter for urethral pressure and function measurement.
- Utilize bench-top testing and an animal model for validation.
Main Methods:
- Bench-top evaluation using a silastic tube with a restriction.
- In vivo testing in anesthetized dogs, measuring urethral skeletal sphincter pressure via electrical stimulation.
- Simultaneous recording of anal sphincter pressure.
Main Results:
- Bench-top tests demonstrated accurate pressure recordings with localized peaks.
- Animal studies showed rapid urethral pressure responses to electrical stimulation.
- Significantly higher peak pressures were recorded at the skeletal urethral sphincter compared to the proximal urethra.
Conclusions:
- Balloon-pressure recordings provide rapid and adequate data.
- The method successfully identified high-pressure contractions specific to the skeletal urethral sphincter in an animal model.
- Balloon-tipped catheters show promise for evaluating detrusor-sphincter dyssynergia and stress urinary incontinence.
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Uroflowmetry is a non-invasive urodynamic test designed to measure various aspects of urination, including volume, flow rate, and the time to void. This test is crucial for diagnosing and assessing conditions such as bladder outlet obstruction, bladder dysfunction, incomplete bladder emptying, incontinence, and urinary tract blockages caused by benign prostatic hyperplasia (BPH) and urethral strictures.Pre-Test Instructions:Before a uroflowmetry test, patients are typically advised to drink...
Measurement of Fluid Pressure
Fluid pressure is commonly measured using devices called manometers, which rely on liquid columns to indicate pressure differences. The height of a liquid column in a manometer reflects the pressure exerted by the fluid, providing a simple yet effective means of measurement. Different types of manometers serve specific purposes based on their configurations and the type of fluids involved.
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Equipments Used To Measure Blood Pressure
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...