Transcatheter closure of moderate-to-large patent ductus arteriosus in infants using Amplatzer duct occluder

Jou-Kou Wang1, Mei-Hwan Wu, Ming-Tai Lin

  • 1Department of Pediatrics, National Taiwan University Hospital, College of Medicine, National Taiwan University, Taipei, Taiwan. jkww@ntuh.gov.tw

Insights

Transcatheter closure of patent ductus arteriosus (PDA) in infants using the Amplatzer duct occluder (ADO) is safe and effective. This minimally invasive procedure shows high success rates with minimal complications for infant PDA treatment.

Area of Science:

  • Pediatric Cardiology
  • Interventional Cardiology
  • Medical Devices

Background:

  • Transcatheter closure of patent ductus arteriosus (PDA) in infants presents significant challenges.
  • Existing methods may have limitations in this vulnerable population.

Purpose of the Study:

  • To evaluate the safety and efficacy of the Amplatzer duct occluder (ADO) for transcatheter closure of PDA in infants.
  • To assess procedural success rates and short-term outcomes.

Main Methods:

  • A cohort of 46 infants underwent PDA closure with the Amplatzer duct occluder (ADO).
  • The ADO was indicated for ductus diameters >= 2.5 or 3 mm, with device size 1-3 mm larger than the ductal diameter.
  • Echocardiographic follow-up was performed at 1 and 3 months post-procedure.

Main Results:

  • Successful deployment of the ADO was achieved in 45 out of 46 patients (97.8% success rate).
  • No severe complications were reported during the procedure.
  • A small residual shunt was observed in 4 patients at 1 month, resolving by 3 months.
  • One patient experienced moderate left ventricular outflow tract obstruction 2.3 years post-procedure.

Conclusions:

  • Transcatheter closure of PDA in infants using the Amplatzer duct occluder (ADO) is a safe and effective therapeutic option.
  • The ADO demonstrates a high success rate and favorable short-term outcomes for infant PDA closure.