Acquiring diagnostic DaTSCAN images in claustrophobic or difficult patients using a 180 degrees configuration
Alp Notghi1, Joseph O'Brien, Elizabeth A Clarke
1City Hospital, Sandwell and West Birmingham Hospitals NHS Trust, Dudley Road, Birmingham, B18 7QH, UK. alp.notghi@swbh.nhs.uk
Background:
In this study, we have investigated the feasibility of a 180 degrees DaTSCAN brain SPECT acquisition. This technique has the advantage of being 'open view' for the patient and therefore more acceptable for claustrophobic patients. It also enables easier access for a technologist to hold the patient's head during acquisition to reduce movement in confused patients or in those with severe tremor.
Methods:
In the first part of this study, we validated the practicality and image quality of a 180 degrees acquisition using a DaTSCAN Alderson head phantom with different camera configurations on GE Infinia and Philips AXIS gamma cameras. The effect on image quality of using half the acquisition time was also assessed. In the second part of the study, 50 sets of patient data were reprocessed by reconstructing half of the 360 degrees data to mimic a single-head 180 degrees acquisition. The 180 degrees images were then compared with 360 degrees images for the same patient using a visual score system. The effect of half-time 180 degrees data acquisition on quantification was also assessed using GE QuantiSPECT software.
Results:
All phantom images from 180 degrees acquisitions contained some degree of distortion at the periphery, but clearly retain the presence of centrally positioned caudate and putamen; hence 180 degrees acquisitions were deemed to produce clinically useful diagnostic images. The shorter (half) acquisition time leads to noisier but acceptable images for all configurations. In the patient study, there was complete agreement between the two reporters with no clinical difference in the diagnostic accuracy between the 180 degrees and 360 degrees images. However, 6 of 50 180 degrees images were marked as poor quality but reportable, compared with 0 of 50 in 360 degrees images. Quantification gave consistently lower nuclei to background ratio values for 180 degrees compared with 360 degrees for normal and abnormal patients.
Conclusion:
It is possible to obtain diagnostic DaTSCAN images using 180 degrees acquisition in difficult patients. The total length of the study may also be shortened by half using L-mode or V-mode camera head configuration.
More Related Videos
10:25A Dorsal Skinfold Window Chamber Tumor Mouse Model for Combined Intravital Microscopy and Magnetic Resonance Imaging in Translational Cancer Research
Published on: April 12, 2024
12:09A Novel High-resolution In vivo Imaging Technique to Study the Dynamic Response of Intracranial Structures to Tumor Growth and Therapeutics
Published on: June 16, 2013
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...