Related Experiment Video
Updated: Jun 17, 2026

The Flexible Rhino-Laryngoscope for Awake Nasotracheal Intubation
Published on: August 2, 2024
Displacement of the epiglottis during intubation with the Pentax-AWS Airway Scope
Akihiro Suzuki1, Norifumi Katsumi, Takashi Honda
1Department of Anesthesiology and Critical Care Medicine, Asahikawa Medical College, Midorigaokahigashi 2-1-1-1, Asahikawa, 078-8510, Japan. masuikasuzuki@yahoo.co.jp
Abstract:
The Pentax-AWS system is a rigid indirect video laryngoscope with integrated tube guidance. Complications associated with this device are not well understood. We report two cases of epiglottis malposition during intubation with the Pentax-AWS. The standard technique of using the Pentax-AWS system involves direct elevation of the epiglottis for exposure of the vocal cords. The blade tip should be passed posterior to the epiglottis for laryngeal exposure, but pressure on the anterior surface of the epiglottis by the tip can rarely happen even during the correct maneuver. Although the Pentax-AWS provides clear images of the airway structures, it is sometimes difficult to observe the epiglottis continuously because the camera is located beneath the blade tip. Consequently, the view of the epiglottis from the camera may be impeded by the blade tip and may result in undiagnosed epiglottis malposition. The AWS's structural feature and its approach to the larynx can be associated with increased chance of unexpected epiglottis folding. It is particularly important to confirm normal position of the epiglottis during withdrawal of the device to prevent this complication.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required