Related Experiment Video
Updated: Jun 17, 2026

Measuring Spray Droplet Size from Agricultural Nozzles Using Laser Diffraction
Published on: September 16, 2016
Evaluation of droplet velocity and size from nasal spray devices using phase Doppler anemometry (PDA)
Xiaofei Liu1, William H Doub, Changning Guo
1U.S. Food & Drug Administration, Division of Pharmaceutical Analysis, 1114 Market Street, Room 1002, St. Louis, MO 63101, United States.
Abstract:
To determine aerosol deposition during the inhalation drug delivery, it is important to understand the combination of velocity and droplet size together. In this study, phase Doppler anemometry (PDA) was used to simultaneously characterize the aerosol velocity and droplet size distribution (DSD) of three nasal spray pumps filled with water. Thirteen sampling positions were located in the horizontal cross-sectional area of the nasal spray plumes at a distance of 3cm from the pump orifice. The results showed droplet velocities near the center of the spray plume were higher and more consistent than those near the edge. The pumps examined showed significant differences in their aerosol velocity at the center of the spray plume, which suggest that this metric might be used as a discriminating parameter for in vitro testing of nasal sprays. Droplet size measurements performed using PDA were compared with results from laser light scattering measurements. The ability of PDA to provide simultaneous measurements of aerosol velocity and size makes it a powerful tool for the detailed investigation of nasal spray plume characteristics.

