Related Experiment Video
Updated: Jun 17, 2026

In Situ Surface Temperature Measurement in a Conveyor Belt Furnace via Inline Infrared Thermography
Published on: May 30, 2020
Uncertainty of a hybrid surface temperature sensor for silicon wafers and comparison with an embedded thermocouple
1Department of Mechanical Engineering, Toyo University, 2100 Kujirai, Kawagoe, Saitama 350-8585, Japan. iuchi@toyonet.toyo.ac.jp
Abstract:
We have developed a user-friendly hybrid surface temperature sensor. The uncertainties of temperature readings associated with this sensor and a thermocouple embedded in a silicon wafer are compared. The expanded uncertainties (k=2) of the hybrid temperature sensor and the embedded thermocouple are 2.11 and 2.37 K, respectively, in the temperature range between 600 and 1000 K. In the present paper, the uncertainty evaluation and the sources of uncertainty are described.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Uncertainty in Measurement: Reading Instruments
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Biasing of Metal-Semiconductor Junctions
In Schottky junctions, where the semiconductor is n-type, applying a positive voltage to the metal relative to the semiconductor reduces its Fermi...