Quantitative calibration of X-ray film response in the 5-keV to 1.3-MeV region
C M Dozier1, J V Gilfrich, L S Birks
1The Naval Research Laboratory, Washington, DC 20390, USA.
Abstract:
Response vs energy was measured for Kodak No-Screen and Rapid-Process films. Monochromatic incident radiation was obtained by fluorescent excitation of the characteristic x-ray lines in an element such as copper followed by crystal diffraction to pass only the Kalpha component. Fluorescing elements from atomic number 23(V) to 60(Nd) covered the energy range 4.95-37.0 keV. A single point at 1.25 MeV was obtained for gamma radiation from (60)Co. A proportional detector measured the incident flux density (photons cm(-2) sec(-1)). The No-Screen film is more sensitive than Rapid-Process film and has greater contrast. At 8 keV, to obtain a photographic density of unity requires 3.76 x 10(7) photons/cm(2) for No-Screen film and 4.19 x 10(7) photons/cm(2) for Rapid-Process film; at 20 keV, No-Screen film requires 7.47 x 10(7) photons/cm(2) and Rapid-Process film requires 12.87 x 10(7) photons/cm(2).
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:23A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
Related Concept Videos
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
X-ray Imaging
Flame Photometry: Lab