Related Experiment Video
Updated: Jun 17, 2026

Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
Shock advisory system for heart rhythm analysis during cardiopulmonary resuscitation using a single ECG input of
Vessela Krasteva1, Irena Jekova, Ivan Dotsinsky
1Centre of Biomedical Engineering Prof. Ivan Daskalov, Bulgarian Academy of Sciences, Acad. G. Bonchev Str. Bl. 105, 1113 Sofia, Bulgaria. vessika@clbme.bas.bg
Abstract:
Minimum "hands-off" intervals during cardiopulmonary resuscitation (CPR) are required to improve the success rate of defibrillation. In support of such life-saving practice, a shock advisory system (SAS) for automatic analysis of the electrocardiogram (ECG) contaminated by chest compression (CC) artefacts is presented. Ease of use for the automated external defibrillators (AEDs) is aimed and therefore only processing of ECG from usual defibrillation pads is required. The proposed SAS relies on assessment of outstanding components of ECG rhythms and CC artefacts in the time and frequency domain. For this purpose, three criteria are introduced to derive quantitative measures of band-pass filtered CC-contaminated ECGs, combined with three more criteria for frequency-band evaluation of reconstructed ECGs (rECG). The rECGs are derived by specific techniques for CC waves similarity assessment and are reproducing to some extent the underlying ECG rhythms. The rhythm classifier embedded in SAS takes a probabilistic decision designed by statistics on the training dataset. Both training and testing are fully performed on real CC-contaminated strips of 10 s extracted from human ECGs of out-of-hospital cardiac arrest interventions. The testing is done on 172 shockable strips (ventricular fibrillations VF), 371 non-shockable strips (NR) and 330 asystoles (ASYS). The achieved sensitivity of 90.1% meets the AHA performance goal for noise-free VF (>90%). The specificity of 88.5% for NR and 83.3% for ASYS are comparable or even better than accuracy reported in literature. It is important to note that, the aim of this SAS is not to recommend shock delivery but to advice the rescuers to "Continue CPR" or to "Stop CPR and Prepare for Shock" thus minimizing "hands-off" intervals.
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiopulmonary Resuscitation I: Adult