Related Experiment Video
Updated: Jun 17, 2026

Robotic Ablation of Atrial Fibrillation
Published on: May 29, 2015
Persistent iatrogenic atrial septal defect after a single-puncture, double-transseptal approach for pulmonary vein
Andreas Rillig1, Udo Meyerfeldt, Markus Kunze
1Department of Cardiology, Schwarzwald-Baar-Klinikum Villingen-Schwenningen, Academic Hospital of the University of Freiburg Vöhrenbacherstr, 23, Villingen-Schwenningen, Germany. arillig5@yahoo.de
Aims:
Persistent iatrogenic atrial septal defect (iASD) after transseptal puncture for pulmonary vein isolation (PVI) has been described recently as a complication of PVI. No data exists evaluating systematically the incidence and clinical implications of iASDs after PVI using a remote robotic navigation system (RNS) with sheaths with a distinct larger outer diameter.
Methods And Results:
In this prospective study, 40 patients with either paroxysmal (n = 22, 55%) or persistent symptomatic atrial fibrillation were treated with circumferential PVI using an RNS. In all patients, a single-puncture, double-transseptal approach was used to access the left atrium. Transoesophageal echocardiography was performed before and the day after PVI as well as after a 3 and 6 months follow-up (FU). The day after ablation an iASD was detected in 38 of 40 (95%) patients with a mean diameter of 3.45 +/- 1.5 mm. At 6-month FU, the iASDs were closed in 30 of 39 (78.9%) patients. During the 6-month FU period, no patient died or suffered from cerebral or cardiac embolism.
Conclusion:
After a single-puncture, double-transseptal approach for PVI using the RNS, iASDs show a high spontaneous closure rate of 78.9% after a 6-month FU period. Persistent iASDs following PVI with the RNS are not associated with an increased rate of paradoxical embolism or with relevant shunting.