Common fetal measurements: a comparison between ultrasound and magnetic resonance imaging
Anagha P Parkar1, Øystein E Olsen, Knut Gjelland
1Department of Radiology, Haukeland University Hospital, Bergen, Norway.
Background:
Ultrasound has been the method of choice for antenatal fetal assessment for the past three decades; however, problems may arise in cases of oligohydramnion, unfavorable position of the fetus, and maternal obesity.
Purpose:
To compare ultrasound (US) and magnetic resonance imaging (MRI) for common fetal measurements at 19-30 weeks' gestation, and to assess the effect of high maternal body-mass index (BMI).
Material And Methods:
59 low-risk singleton pregnancies were enrolled in a prospective blinded cross-sectional study. In a first session, an experienced obstetrician used a high-resolution US technique and in a second session on the same day MRI was used to measure biparietal diameter (BPD), head circumference (HC), mean abdominal diameter (MAD), abdominal circumference (AC), and femur length (FL). Inter- and intraobserver and intermodality variability was determined using Bland-Altman plots. The effect of maternal BMI was assessed using Spearman's statistics.
Results:
A total of 45 women aged 19-43 years (median 29 years) attended both US and MRI at median 22 weeks' gestation. The mean differences between US and MRI were 1.6 mm for HC (95% confidence interval [CI] -1.0, 4.3 mm), 1 mm for AC (95% CI -0.2, 4.0 mm), 0.2 mm for MAD (95% CI -0.7, 1.2 mm), 2.2 mm for BPD (95% CI 1.7, 2.7 mm), and 4.6 mm for FL (95% CI 2.9, 6.4 mm). Maternal BMI did not affect the results (Spearman' rho 0.054-0.277; P=NS). The intraobserver agreement for all MRI measurements was acceptable, except for FL, while the interobserver agreement was poor.
Conclusion:
There was good agreement between US and MRI for common fetal measurements, but not for all (i.e., BPD and particularly FL). MRI had a poor interobserver agreement, underscoring the need for technical refinement and reference ranges specifically established for MRI.