Microemboli in our bypass circuits: a contemporary audit.
Timothy W Willcox1, Simon J Mitchell
1Green Lane Clinical Perfusion, Auckland City Hospital, New Zealand, and Faculty of Medicine and Health Sciences, University of Auckland, Auckland, New Zealand.
Cardiopulmonary bypass (CPB) circuits generate microemboli, primarily from left ventricular venting and air entrainment. An Emboli Detection and Classification (EDAC) audit identified key sources, aiding CPB safety improvements.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Medical Device Technology
Background:
- Cardiopulmonary bypass (CPB) circuits can introduce arterial microemboli.
- Clinical studies on CPB-related microemboli are limited, with most research being in vitro.
- Identifying microemboli sources is crucial for improving CPB safety and patient outcomes.
Purpose of the Study:
- To prospectively audit microemboli generation within a contemporary CPB circuit using the Emboli Detection and Classification (EDAC) system.
- To quantify emboli number and volume at different points in the CPB circuit.
- To identify the primary sources of microemboli during clinical CPB procedures.
Main Methods:
- A prospective clinical audit was conducted on standard CPB circuits in 12 patients.
- Three EDAC system probes were placed on the venous line, hard-shell venous reservoir (HSVR) outlet, and arterial line filter.
- Emboli counts and volumes were recorded at 30-second intervals, synchronized with the perfusion data management system.
Main Results:
- Microemboli counts were significantly higher downstream of the HSVR than the venous line.
- The arterial line filter significantly reduced both emboli count and volume.
- Microemboli generation increased with the left ventricular vent in operation, with air entrainment from the venous line being a major source.
Conclusions:
- The EDAC system effectively quantifies microemboli in CPB circuits.
- Left ventricular venting and venous line air entrainment are significant sources of microemboli.
- Continuous auditing and potential component redesign can enhance CPB safety and improve patient outcomes.