A clinical comparison of bubble elimination in Quadrox and Polystan oxygenators
Mario Jirschik1, Cornelius Keyl, Friedhelm Beyersdorf
1Department of Cardiovascular Surgery, Universitaeres Herzzentrum Freiburg-Bad Krozingen, Germany. mario.jirschik@herzzentrum.de
Background:
Microbubbles generated during heart surgery on extracorporeal circulation have been implicated as a possible cause of postoperative neurocognitive dysfunction and negative outcome. The main sources of microbubbles in the extracorporeal circuit are air leaking from the venous cannulation site, air delivered by drug and volume administration, during the taking of blood samples, during hemofiltration, and by using vacuum-assisted venous drainage (VAVD). Membrane oxygenators, although not designed for the elimination of gaseous microbubbles, can eliminate much of this air. AIM OF THE STUDY / HYPOTHESIS: The aim of this study was to test the hypothesis that the amount of microbubbles passing through the oxygenator varies depending on the oxygenator design.
Methods:
one hundred patients undergoing aortic valve replacement with cardiopulmonary bypass (CPB) were included in a retrospective clinical trial assessing the bubble elimination of the two oxygenators routinely used in our clinic. The oxygenators (Quadrox-i Adult / Polystan Safe Maxi) are manufactured with the same hollow-fiber material, but display different designs. Bubbles were detected by a two-channel ultrasound bubble counter. The probes were placed directly at the inlet and the outlet of the oxygenators. The filtration index was calculated by the bubble counter as a measure of the percentage reduction of bubble volume and number.
Results:
The Quadrox has a significantly higher filtration index (92.3% [89.7; 95.1], median [25(th),75(th) percentile]) than the Polystan (74.9% [64.2;80.9], p<0.001), indicating the superior ability of the Quadrox to eliminate microbubbles.
Conclusion:
Our study demonstrates major differences in the capacity of bubble elimination of the oxygenators being investigated, possibly depending on oxygenator design.
More Related Videos
13:17Manufacture of Concentrated, Lipid-based Oxygen Microbubble Emulsions by High Shear Homogenization and Serial Concentration
Published on: May 26, 2014
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Transport in the Blood
Bioreactor Controls-II
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications