Adaptive force sonorheometry for assessment of whole blood coagulation
F William Mauldin1, Francesco Viola, Theresa C Hamer
1Department of Biomedical Engineering, University of Virginia, Charlottesville, VA 22908, USA.
Background:
Viscoelastic diagnostics that monitor the hemostatic function of whole blood (WB), such as thromboelastography, have been developed with demonstrated clinical utility. By measuring the cumulative effects of all components of hemostasis, viscoelastic diagnostics have circumvented many of the challenges associated with more common tests of blood coagulation.
Methods:
We describe a new technology, called sonorheometry, that adaptively applies acoustic radiation force to assess coagulation function in WB. The repeatability (precision) of coagulation parameters was assessed using citrated WB samples. A reference range of coagulation parameters, along with corresponding measurements from prothrombin time (PT) and partial thromboplastin time (PTT), were obtained from WB samples of 20 healthy volunteers. In another study, sonorheometry monitored anticoagulation with heparin (0-5 IU/ml) and reversal from varied dosages of protamine (0-10 IU/ml) in heparinized WB (2 IU/ml).
Results:
Sonorheometry exhibited low CVs for parameters: clot initiation time (TC1), <7%; clot stabilization time (TC2), <6.5%; and clotting angle (theta), <3.5%. Good correlation was observed between clotting times, TC1 and TC2, and PTT (r=0.65 and 0.74 respectively; n=18). Linearity to heparin dosage was observed with average linearity r>0.98 for all coagulation parameters. We observed maximum reversal of heparin anticoagulation at protamine to heparin ratios of 1.4:1 from TC1 (P=0.6) and 1.2:1 from theta (P=0.55).
Conclusions:
Sonorheometry is a non-contact method for precise assessment of WB coagulation.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Coagulation
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...