Multi-site, multi-country evaluation of analytical and operational performance of a low-mid volume chemiluminescent
M H De Keijzer1, S Perkins, V Motta
1Laboratorium en Trombosedienst, Ziekenhuis Rivierenland, Tiel, Netherlands.
Background:
A new automated immunoassay low-mid volume (< or = 250 immunoassays/day) chemiluminescent analyzer, Abbott Architect i1000sR, was evaluated by seven laboratories around the world (4 in Europe, one each in Canada, Japan, and the U.S.A.) to demonstrate equivalent performance for key operating characteristics (e.g., precision, turn around time, limit of detection, functional sensitivity, and linearity).
Methods:
The laboratories followed standard protocols to assess precision, limit of detection (LoD), functional sensitivity, assay linearity, method comparison, and sample carryover. Turn around time for three stat assays (beta-hCG, BNP, and CK-MB) and the time required to complete workloads of 50 and 100 tests with a mixture of 75% routine tests and 25% stat tests was also evaluated.
Results:
Total precision was typically < 5% CV for nine immunoassays. Analytical performance met design goals and demonstrated equivalency to package insert data for assays on market and in use for an existing high volume immunoassay system. Stat turn around times were consistent with the fixed analytical time of 15.6 minutes and met the expectations of the laboratories. Measured test throughput ranged from 47 - 54 tests per hour and demonstrated that the analyzer was fit for the intended purpose of supporting a laboratory that performs < or = 250 immunoassays per day.
Conclusions:
A multisite, international analyzer familiarization study is a practical means of confirming that a new instrument meets both a manufacturer's design specifications and users' real world expectations and provides a pragmatic test for the system. The experience of investigators at seven sites around the world indicates that a new fully automated chemiluminescent system is suitable for use.