Aims:
Cardiac output (CO) measurements from three-dimensional (3D) trans-mitral Doppler echocardiography are prone to error as manual selection of the region of interest (i.e. the site of measurement) is required. We newly developed an automated, user-independent algorithm to select the site of colour Doppler CO measurement. We aimed to validate this new method by benchmarking it against thermodilution, the current gold standard for CO measurements.
Methods And Results:
Transoesophageal colour 3D Doppler echocardiographic studies were obtained from 15 patients who also had received a pulmonary catheter for invasive CO measurements. Trans-mitral flow was determined using a novel operator-independent algorithm to automatically select the optimal site of measurement. The operator-independent CO measurements were referenced against thermodilution. A good correlation was found between operator-independent Doppler flow computations and thermodilution with a mean bias of 0.09 L/min, standard deviation of bias 1.3 L/min, and a 26% error (2 SD/mean CO). Mean CO was 4.94 L/min (range 3.10-7.10 L/min).
Conclusion:
Our findings demonstrate that CO computation from transoesophageal colour 3D Doppler echo can be automated concerning the site of velocity measurement. Our operator-independent algorithm provides an objective and reproducible alternative to thermodilution.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...